8.
Lift and lower the movable plate several times to
make sure that there is no binding, etc.
9.
Secure the sample probe motor's wire leads with
the two cable clamps [E] and the three cable clips
[F].
Make sure that there is a service loop [L] between
the body of the motor and the first cable clamp.
Replacing the osmometer module
Refer to
Removing the osmometer module
and callouts.
1.
Slide the osmometer module into the instrument.
2.
Secure the osmometer module to the mounting
bracket using the four socket head cap screws and
washers [D].
Replacing the power supply and controls
assembly
Refer to
assembly
1.
2.
3.
4.
L
Reconnecting the control PCB
1.
2.
3.
for images
4.
5.
6.
Replacement Procedure
OsmoTECH® PRO Sample Probe Motor
Removing the power supply and controls
for images and callouts.
Slide the power supply and controls assembly into
the base assembly.
Secure the power supply and controls assembly to
the rear of the chassis with the two screws [C].
Place the instrument on its back side.
Secure the power supply and controls assembly to
the bottom of the chassis with the two screws [B].
Place the instrument in an upright position.
Plug the A1, A2, A3, and A4 connectors on the
cooling system harness into their mating
connectors.
Plug the B1, B2, B3, B4, B6, B7, B8, B9, and B10
connectors on the motor/sensor cable harness into
their mating connectors.
Plug the C10 and C11 connectors into the control
PCB.
Secure the C10 sample probe cable and the C11
block probe cable to the side of the power supply
and controls assembly with the cable clamp and
screw [A].
Reassemble the instrument by performing the
reassembly instructions in OsmoTECH PRO
Disassembly and Reassembly.
Page 4 222104PM Rev 0