Andatech ALX3000 Benutzerhandbuch - Seite 8

Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Prüfgeräte Andatech ALX3000 herunter. Andatech ALX3000 20 Seiten. Ignition interlock

Andatech ALX3000 Benutzerhandbuch
MEASUREMENTS & CONDITIONS FOR TEST
Use a clean mouthpiece for each test. Mouthpiece should be replaced after each positive BAC
test result or when it is dirty.
Do not allow samples containing smoke, tobacco or other chemical products to enter the unit,
as damage to the fuel cell can occur. Cigarette smoke may also increase the blood alcohol
concentration value displayed. Allow one minute after smoking before testing.
There must be an interval of at least 15 minutes after alcohol consumption as alcohol residue
in the mouth may give a false reading. Such residue may also be left by aromatic drinks,
mouthwash, medicines and drops, as well as by belching and vomiting. Rinsing your mouth
with water or non-alcoholic drinks does not substitute for an interval.
A person's blood alcohol concentration (BAC) can continue to rise for up to 2 hours after
drinking alcohol. If a result is close to the designated BAC, it is recommended to wait and
retest after 20 minutes.
It can take 10 hours or more for a person's BAC to return to zero after a high BAC has been
reached.
LIABILITY FOR PROPER FUNCTION OR DAMAGE
The liability for proper functioning of the instrument is irrevocably transferred to the owner
or operator in the event the instrument is serviced or repaired by personnel not employed or
authorised by Andatech. In such an event, the warranty will be void.
WARRANTY
The manufacturer warrants for 1 year
from the date of purchase, the product to
be free from defects in workmanship or
material (excluding recalibration) under
normal use. The manufacturer's obligation
under this warranty is limited to replacing,
adjusting or repairing the unit if returned
along with proof of purchase. This warranty
is void if the unit has been tampered with,
damaged or abused in any way.
Visit my.andatech.com.au to register your
warranty online and receive calibration
reminders.
8
CALIBRATION AND SERVICE
Andatech recommends the Andatech
ALX3000 to be calibrated every 12 months
to maintain its accuracy. The Andatech
ALX3000 cannot be used to take a test
once it has passed its calibration due date.
To register your product warranty,
receive calibration reminders and
book in a calibration, please go to
my.andatech.com.au or contact:
Andatech
Hotline: 1300 800 200
Website: www.andatech.com.au
®
®