ANDERSON-NEGELE MPP Handbuch für Installation und Inbetriebnahme - Seite 3
Blättern Sie online oder laden Sie pdf Handbuch für Installation und Inbetriebnahme für Transmitter ANDERSON-NEGELE MPP herunter. ANDERSON-NEGELE MPP 4 Seiten. Modular pressure transmitter
Output: 4Ma – 20Ma or 20mA – 4mA (HART only)
1.
Press "M" – Sensor displays "COnF"
2.
Press "E" – Sensor displays "PSI" or "BAR"
3.
Press "M" – Sensor displays "4-20" or "20-4"
4.
Press "U" or "D" – Sensor toggles between "4-20" or "20-4"
5.
If no other operation is to be accessed exit to "RUN" mode via pressing "M" repeatedly until process value is displayed
PRESSURE RANGE
LRV: Lower range value
Sensors equipped with "Compound" style measurement cells (stem) may configure the LRV. "Absolute" stems are predefined as 0
and are not reconfigurable
1.
Press "M" – Sensor displays "COnF"
2.
Press "E" – Sensor displays "PSI" or "BAR"
3.
Press "M" – Sensor displays "4-20" or "20-4" (HART only)
4.
Press "M" – Sensor displays "LRV"
5.
Press "E" – Sensor displays present LRV value – Example: "0"
6.
Press "U" or "D" to set desired LRV value– Sensor display increases or decreases accordingly
Note: continue to URV
URV: Upper range value
Note: starting from LRV above
7.
Press "M" – Sensor displays "URV"
8.
Press "E" – Sensor displays present URV value – Example: "50"
9.
Press "U" or "D" to set desired URV value – Sensor display increases or decreases accordingly
10.
Press "E" to store displayed value – Sensor displays "URV"
11.
If no other operation is to be accessed exit to "RUN" mode via pressing "M" repeatedly until process value is displayed
Output Damping
1.
Press "M" – Sensor displays "COnF"
2.
Press "E" – Sensor displays "PSI" or "BAR"
3.
Press "M" – Sensor displays "4-20" or "20-4" (HART only)
4.
Press "M" – Sensor displays "LRV"
5.
Press "M" – Sensor displays "URV"
6.
Press "M" – Sensor displays "dMPg"
7.
Press "E" – Sensor displays current damping value from 0 - 10
8.
Press "U" or "D" to display desired damping value– Sensor display increases or decreases accordingly
9.
Press "E" to store displayed value – Sensor displays "dMPg"
10. If no other operation is to be accessed exit to "RUN" mode via pressing "M" repeatedly until process value is displayed
Setting Display Process Variable - Functionality in display interface only.
For additional display interface functionalities see section 7.2 of the manual.
The following Engineering units may be selected:
Millibar, Torr & Kg/cm2, PSI, BAR, kPA, In H2O, In Hg, mm H2O, mA
1.
Press "M" – Sensor displays "CONF"
2.
Press "E" – Sensor displays "PSIG" or "BARG"
3.
Press "M" – Sensor displays "4-20" or "20-4"(HART only)
4.
Press "M" – Sensor displays "LRV"
5.
Press "M" – Sensor displays "URV"
6.
Press "M" – Sensor displays "DAMP"
7.
Press "M" – Sensor displays "ALRM1"
8.
Press "M" – Sensor displays "ALRM2"
9.
Press "M" – Sensor displays currently set engineering unit
10. Press "U" or "D" repeatedly to select the desired engineering unit
11. Press "M" to proceed to next operation or repeatedly to exit to "run" mode
Factory configuration reset
Perform if a return to the original factory configuration is desired
1.
Press "M" – Sensor displays "COnF"
2.
Press "E" – Sensor displays "PSI" or "BAR"
3.
Press "M" – Sensor displays "4-20" or "20-4" (HART only)
4.
Press "M" – Sensor displays "LRV"
5.
Press "M" – Sensor displays "URV"
6.
Press "M" – Sensor displays "dMPg"
7.
Press "M" – Sensor display "FrES"
8.
Press "E" – Sensor display "nO"
9.
Press "U" or "D" to display "YES"
10. Press "E" – Sensor display "FrES"
11. exit to "RUN" mode via pressing "M" repeatedly until process value is displayed