Andon KD-5904 Betriebshandbuch - Seite 18

Blättern Sie online oder laden Sie pdf Betriebshandbuch für Blutdruckmessgerät Andon KD-5904 herunter. Andon KD-5904 19 Seiten.

DN:KD-5904-SMSY01 V3.0
Radiated RF
3 V/m 80 MHz to 2.5
IEC 61000-4-3
GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the KD-5904 is used
exceeds the applicable RF compliance level above, the KD-5904 should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the KD-5904.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the KD-5904
The KD-5904 is intended for use in an electromagnetic environment in which radiated RF
d
3 V/m
d
GHz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey,
should be less than the
compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment
marked with the following
symbol:
Table 4
18/19
Operation Guide
1
2 .
P
80 MHz to 800
MHz
2
3 .
P
800 MHz to 2,5
b
JS-KD5904-03-V2.0
a