Droper Field 1000 Betriebsanleitung - Seite 5

Blättern Sie online oder laden Sie pdf Betriebsanleitung für Medizinische Ausrüstung Droper Field 1000 herunter. Droper Field 1000 16 Seiten. Mechanical pressure infuser
Auch für Droper Field 1000: Gebrauchsanweisung Handbuch (15 seiten)

2.
Without releasing the pressure exerted on the cover, the operator
then exerts upwards pressure on the two disarming side levers
by pinching his/her middle fingers of each hand upwards and
inwards. This movement will result in pivoting the closing spring
clips inwards and releasing the cover.
2.2
Inserting the infusion line
With outstretched arms, the operator places his/her middle finger
of each hand underneath the disarming side lever. The device opens
in three steps:
1. The operator firstly exerts pressure in a downwards direction on
the cover using his/her body weight and with outstretched arms.
This action places pressure on the mobile plate. The resulting
small downwards movement can be observed.
3.
While holding the two open levers in position, release
the pressure exerted on the cover. Then open the
cover by pushing it upwards using an index finger.
Where necessary, install an extension with a 3-way
stopcock at the start of the line to be able to check the
correct operation of the device during the infusion.
Check that the choke or flow regulator is set to the OFF
position.
Check that the air intake valve in the drip chamber is in
the closed position.
Insert the piercing pin at the bottom into the tubing port
of the infusion bag.
5 / 16