Droplet SP2 Handbuch - Seite 5
Blättern Sie online oder laden Sie pdf Handbuch für Messgeräte Droplet SP2 herunter. Droplet SP2 13 Seiten. Beam-scan camera and software
Manual, SP2 Beam Scan Camera and Software
The following is the listing of the mightex.prop file:
; gain is normally 1.0, but can be as high as 16
gain=1.0
; gamma is from 0.0 to 2.0
gamma=1.0
; contrast is from 0 to 100 in percent
contrast=50
; brightness is from 0 to 100 in percent
brightness=50
; exposure time is in microseconds, so 30000 us = 30 ms
; can be up to 750 ms (750000)
exposure time=10000
;
; the frame rate slows greatly at the higher exposure times. 750000
; is the maximum exposure time on the Mightex cameras. That should
; be enough to integrate multiple laser pulses if needed.
It is recommended that exposure time be lengthened until saturation happens. Then back off
the exposure time by 500 or 1000 units. Saturation will not cause damage to the camera, and it
can be useful to operate in this mode, as the base of the beam will be easier to visualize. No
studies have been done on modifying the other parameters in the mightex.prop file.
CAUTION: The Mightex camera is fitted with a 5 OD filter for standard operation. This reduces
the laser power to a range where the camera can provide a good beam profile. The laser power
that is available at the YAG power monitor can be up to 200mW. The Mightex camera is rated to
withstand laser power of up to a MAXIMUM of 70mW only under certain conditions, which are
unspecified. Operation of the camera on the SP2 with any neutral density filter having an OD of
less than 2 will void the warranty of the product. The Measured leakage should be between 80
– 100 mw
5
DOC-0175 Rev D
© 2017 DROPLET MEASUREMENT TECHNOLOGIES