System Specifications

System Specifications
Table 1 lists the specifications of the CEQ System. Use this information to prepare the
environment where the system will be located.
The CEQ System is designed for indoor use only.
CAUTION
Table 1:
System Specifications
Item
Weight (CEQ)
Power Requirements
PC (typical) 6.0A, 50/60Hz
Monitor (17" typical) 1.8A, 50/60Hz
Fuses (CEQ)
Maximum Heat Dissipation
(to room)
PC and Monitor 936W (3194 BTUs/Hour)
Ambient Operating
Temperature Range
Humidity Restrictions
Altitude Restrictions
PC (Tower)
Minimum Specifications
Installation Category
Pollution Degree
Laser Category
4
70.5Kg (155lbs)
CAUTION
CEQ 100-240VAC, 5.0A, 50/60Hz
6.3A Time Delay; 20mm (2 ea.); 200-240VAC
CEQ 400W (1365 BTUs/Hour)
15-30°C (59-86°F)
<85% (non-condensing) @ 30°C (86°F)
up to 2000m (6,562ft)
Beckman certified controller. For specific features, contact your local
Beckman Coulter, Inc. Sales and Service office.
Category II
2
Class 1
WARNING
CEQ™ 8000 Genetic Analysis System
Description
The supply voltage must not exceed 10% of
nominal.
This instrument uses a "Class 3B" laser. The
laser is housed in a sealed container. During
normal operation of the system, laser light is not
accessible to the user. Therefore, overall laser
classification of the sequencer is "Class 1."