6. Detach the wires from the terminal block.
Locate the cables that extends from the Level Sensor and cut any cable ties that bundle it with other cables.
Note: This cable should contain 3 wires #3 Red, #4 Black, #I1 white which attach to a terminal block.
Detach these from the terminal block by inserting and pushing the tip of a flat head screwdriver into the square
Slot on the terminal block that is next to the wire you desire to remove.
This will release the wire. Note the location from which they were removed.
ANKOM Technology • 2052 O'Neil Road • Email: [email protected]• Phone: 315-986-8090 • Fax: 315-986-8091
Replacing X59 Analog level sensor
and Fiber optic cable with glass tip
XT15/XT15I
Service Procedure
X5S012
JAD 4/2/2019
pg. 2 of 4