Ankom XT15 Reinigung

Blättern Sie online oder laden Sie pdf Reinigung für Laborausrüstung Ankom XT15 herunter. Ankom XT15 4 Seiten. Extractor sight glass assembly
Auch für Ankom XT15: Serviceverfahren (5 seiten)

Ankom XT15 Reinigung
1.
Remove the back cabinet.
a. Remove the eight screws (four on each side of
the instrument) that secure the back cabinet
onto the instrument using a 1/8" Allen wrench.
Set aside the screws for later use.
b. Lift up then pull back so the handles on each
side slide out of the notched groove.
2.
Unscrew the compression fitting using a 9/16 wrench.
ANKOM Technology • 2052 O'Neil Road • Email: [email protected] • Phone: 315-986-8090 • Fax: 315-986-8091
Solvent filter cleaning
XT10/XT10I and XT15/XT15I
Service Procedure
Screws
Notched
groove
Figure 3
XTS016
JAD 1/19/2018
Figure 1
Figure 2
pg. 1 of 4