Operator's Manual
The BECSys2 requires no maintenance other than a periodic calibration check and sensor cleaning.
C - 1: Potentiometric Sensors (pH and ORP)
C - 1.1: Electrode Cleaning:
Slow response time and large offsets may indicate the
electrode has become coated. The nature of the
coating will dictate the type of cleaning technique
that should be used.
You may lightly blot the water on a pH sensor
tip on a paper towel, but never vigorously rub
or wipe the pH bulb because this may scratch
the delicate outer layer on the pH glass
impairing its response.
Soft coatings, like bacterial films, are best removed
using a squirt bottle or the water jet from a faucet.
If this is not successful, then gently wipe with a
soft wet cloth.
For a more severe coating, first try a strong
detergent (something similar to Dawn liquid
detergent) and warm water, using a soft brush (like
a toothbrush). Isopropyl alcohol on a Q-tip is
another good choice. Rinse the measuring end in
distilled water before reinstallation.
Greasy and oily coatings are best removed with a
detergent solution or a solvent that will not attack
the sensor body. Methanol and isopropyl alcohol
are good choices for solvents. Acetone, MEK,
THF, or trichloroethane will irreparably harm the
electrode.
Hard coatings, like calcium or lime scale, are best
removed with a solvent appropriate for the
particular coating. A 5% solution of hydrochloric
acid (HCl) would be a good choice for calcium
scale. If unsure of the proper solvent to remove a
hard mineral coating, then alternate between a 5%
hydrochloric acid and a 4% sodium hydroxide
(NaOH) for 10 minutes each. After treating the
electrode with these strong acids or bases, rinse the
electrode with water and soak it in a pH 4 buffer
for at least 1/2 hour.
The platinum tip of an ORP sensor can be cleaned
with an abrasive as a last resort. Gently scour the
platinum with a 600 grit wet emery cloth, or
preferably, a 1-3 micron alumina polishing powder.
Page 4
9487 Dielman Rock Island Industrial Dr. St. Louis, MO 63132 Tel:(314) 567-0088
Section C: Maintenance
C - 1.2: Long-Term Storage:
Save the wetting cap that came with the sensor for
long-term storage.
from the flow-cell, clean it as in routine
maintenance, then store it in the wetting cap using
a pH 4 buffer saturated with potassium chloride
(KCl).
electrolyte from leaching out of the sensors
reference cell. The wetting cap only needs to be
half full. If a number of sites are going to be
serviced, for example, at the end of a season, then
it might be a good idea to carry a pint of 4.0/KCl
storage solution.
After removing the sensor
The potassium chloride will prevent
TECHNOLOGY