4.
Put the screws and spring washers into the screw holes of the flange and pre-tighten 4 turns.
Alternately tighten each diagonally opposite pair of screws to a torque of 1.8 Nm. Ensure
screws and washers are clean before installing. The screws and washers cannot be used if
they are damaged.
3.4 Verification
Please verify sensor operation after reassembly.
4. CONTACT DETAILS
4.1 Druck China Office
Location: 1, Hua Tuo Rd., Zhangjiang Hi Tech Park, Shanghai, 201203
Tel: +86 021 3858 2100, + 86 800 915 9966
Website: www.druck.com/cn
WeChat: BHDS_Druck
4.2 Services and Support Locations
Changzhou China
Tel: +86 400 818 1099
E-mail: [email protected]
English–UNIK5600/5700 Demountable Backend Exchange Manual | 3