Inserting a Clean‑Trace ATP Test UXC or H2O into the
Clean‑Trace Luminometer LX25
• Press the rocker cap to open the test chamber (1) .
• Insert an activated Clean-Trace ATP Test UXC or H2O (2) .
• The Test handle should be visible at the top of the
chamber (3) .
• After the reading is complete, press the rocker cap to open
the test chamber and remove the ATP Test UXC or H2O .
• If an ATP Test UXC or H2O remains in the Clean-Trace
Luminometer LX25 chamber for more than 30 seconds,
the Swab Sensor notifies with an audible tone .
• Keep the Clean‑Trace Luminometer LX25 upright
whenever a test device is in the chamber, to ensure
accurate results and avoid spillage of reagents, which
could damage the sensitive components of the instrument .
• If the Clean-Trace Luminometer LX25 is held at greater
than 68 degrees from vertical when a test is in the
chamber, a message notifies the user that the Clean-Trace
Luminometer LX25 is in the incorrect position for testing
and to use a more vertical position when testing .
Tap "OK" then hold the Clean-Trace
o
Luminometer LX25 in a vertical position and repeat
the measurement .
How to Perform a Quick Test
A quick test is performed under circumstances where results are
not uploaded to the Quality Control Data Manager (e .g . training,
demonstrations, no access to Quality Control Data Manager) .
To perform a Quick Test
• Power on the Clean-Trace
Luminometer LX25 .
• After successful completion of
System Diagnostics this landing
screen with appear:
• Log in is not required to perform a
Quick Test .
• Tap "Quick Test"
• Follow the Clean-Trace ATP Test
UXC or H2O Instructions for Use to
collect the sample, activate the ATP
Test and insert into the Clean-Trace
Luminometer LX25 .
• Tap "Start" to take a measurement .
• Remove the ATP Test UXC or H2O after the measurement
is complete .
• All results should be recorded .
Only the last 20 Quick Tests are saved on the
o
Clean-Trace Luminometer LX25 .
The results of a quick test are not saved to the Quality
o
Control Data Manager .
Communicating with the
3M Quality Control Data Manger (QCDM)
Before the Clean-Trace Luminometer LX25 can be used with
the Quality Control Data Manager (QCDM) it must first be
synchronized to the facility's QCDM organization .
To register the Clean-Trace Luminometer LX25, the facility
account must be linked to the facility's organization and
the 3M™ Synchronization Manager Software installed on a
workstation (desktop/laptop) . Registration cannot be performed
using the Clean-Trace Luminometer LX25 Wi-Fi function .
Register a Clean‑Trace Luminometer LX25
to an Organization on QCDM
1 . Find the Synchronization Manager icon in system tray,
right click on the icon and choose "Show window" .
2 . Power up the Clean-Trace Luminometer LX25 and launch
the app if not already launched .
a . Follow the steps in the set-up
menus for language, date, time
and time zone .
b . Skip Wi-Fi set up .
c . Wait for diagnostics
to complete .
d . When the landing screen
appears -
STOP – remain on
landing screen – DO NOT
LOG IN.
3 . Using the USB cable provided, connect the Clean-Trace
Luminometer LX25 to the workstation running the
Synchronization Manager software .
4 . After a short period, a red, spinning
circle appears on the Clean-Trace
Luminometer LX25 screen,
indicating your Clean-Trace
Luminometer LX25 is connected
to the Synchronization
Manager Software .
7