This device is intended to be operated by properly trained personnel who have carefully read the
operating manual and are familiar with the function of the device. [Refer to the clinical laboratory
method specified by the specimen receptacle manufacturer or established by the medical technology
for the products applications.]
WARNING: For the safety of both the operator and service personnel, care should be taken when using
this centrifuge if handling substances that are known to be toxic, radioactive, or contaminated with
pathogenic microorganisms. Use appropriate personal protection equipment (PPE). When Risk Group II
materials are used, (as identified in the World Health Organization "Laboratory Bio-Safety Manual"), a Bio-
Seal should be employed. In the event that materials of a higher risk group are being used, more than one
level of protection must be provided. The use of flammable or explosive materials as well as those
materials which have a vigorous chemical reaction is prohibited.
Unplug the centrifuge before cleaning or performing maintenance.
WARNING: Inspect centrifuge for cracks or physical damage to cabinet, lid, rotor, or tube holders. Damage
may result in unsafe operation. Discontinue use until repairs have been performed.
This equipment generates, uses, and can radiate radio frequency energy, and if not installed and used in
accordance with this operator manual, may cause interference to radio communications.
Operation of this equipment in a residential area may cause interference, in which case the user will be
required to correct the interference at his own expense.
Important Operating Instructions: This model centrifuge requires at least 2 tube holders inserted when
attempting to run. Running the centrifuge with an empty rotor will cause a motor error and will require a
power cycle to fully reset (see Troubleshooting section on page 9).
Drucker Diagnostics – Customer Service: +1-814-692-7661 –
S
TATEMENTS
| Page 3