ROUBLESHOOTING
Note: The latch must be turned completely clockwise to its stop position
in order for the centrifuge to operate.
The rotor does not spin
freely.
Excessive noise when
the machine is
running.
The centrifuge does
not run.
Drucker Diagnostics +1-814-692-7761 - [email protected] Page | 10
Make sure nothing has fallen into
o
the rotor chamber.
If there is nothing obstructing the
o
rotor, the rotor may be damaged.
contact Drucker Diagnostics for
further assistance.
Check to see that the load is
o
balanced.
Make sure that nothing has fallen
o
into the rotor chamber.
Make sure that the nut in the
o
center of the rotor is tight.
Have a technician test the motor
o
and replace it if necessary.
Check the electrical outlet.
o
Make sure the lid latch is turned
o
completely clockwise to its stop
position.
Check the circuit breaker switch at
o
the bottom left of the machine. If
the switch is white, the breaker has
tripped. Contact Drucker
Diagnostics for further assistance.
The printed circuit board may be
o
damaged. Have a technician test
and replace the circuit board if
necessary.