DS Maref DVT-2600 Service-Handbuch - Seite 12
Blättern Sie online oder laden Sie pdf Service-Handbuch für Medizinische Ausrüstung DS Maref DVT-2600 herunter. DS Maref DVT-2600 16 Seiten.
4
MODE 4
5
MODE 5
6
MODE 6
7
MODE 7
- Check the current voltage of battery ·
maximum voltage : 12.6V±0.2V
- The battery's voltage marking is
below
, connect AC power and
recharge the battery completely..
- Pressure calibration can be operated.
- Start pumping with Manual Pump
manually and check the pressure value
from DVT machine and pressure gauge.
If there is difference between value on
DVT machine and value on Pressure
gauge, check how it different. If the
difference is more than ±15%, then
activate following process. Press the
LEG button to control pressure range
between 40, 80, 120. It will go up
accordingly as the LEG button pressed.
Then Keep doing this process until
make two values are equal. When two
values reach to equal value, press the
INTERVAL button to save the current
value. If error range is more than
10mmHg, the settlement won't be
saved in the program.
(Sensor shoule be replaced if the error
range is too high.)
- All icons are displayed on LCD
- Check the break and unclear display
of letters or icons
- Check the condition of Solenoid valve
- The left number indicates the
sequence of solenoid valve.
- The right number indicates the
discharge voltage of solenoid valves