3M Unitek Ortholux Luminous Curing Ligh Gebrauchsanweisung Handbuch - Seite 16
Blättern Sie online oder laden Sie pdf Gebrauchsanweisung Handbuch für Medizinische Ausrüstung 3M Unitek Ortholux Luminous Curing Ligh herunter. 3M Unitek Ortholux Luminous Curing Ligh 16 Seiten. Fast curing cordless led light
Recommended separation distances between portable and mobile RF
communications equipment and the Ortholux™ Luminous Curing Light unit
The [EQUIPMENT or SYSTEM] is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Ortholux™ Luminous Curing Light unit can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Ortholux Luminous Curing Light unit as recommended
below, according to the maximum output power of the communications equipment
Rated maximum
output of
transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Separation distance according to frequency of transm itter
150 kHz to 80 MHz
0,04
0,11
0,35
1,11
3,50
m
80 MHz to 800 MHz
0,12
0,38
1,20
3,79
12,00
800 MHz to 2,5 GHz
0,40
1,26
4,00
12,65
40,00