DSC ELECTRONICS DP-M Series Benutzerhandbuch - Seite 10
Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Stromversorgung DSC ELECTRONICS DP-M Series herunter. DSC ELECTRONICS DP-M Series 16 Seiten. High power laboratorypowersupplies
DP-M Series User's Manual
Version 12.11.2021
5.5 OVP and OCP Protection Limits
The over voltage and over current protection limits can be adjusted to meet your requirements. By default, these are
set to the maximum value.
1. Press the OV-set or OI-set key to enter OVP or OCP limit setup mode.
2. Tune the adjustment knob to set an OVP or OCP limit value and press OK to confirm.
If the output value exceeds either the set OVP or OCP value, the output is shut down immediately and the ERR icon
along with the OC or OV icon appears. To disable the alarm press the OK button once.
5.6 Front Panel LOCK
To lock the front panel buttons press the LOCK button once. All buttons and the rotary knob are now disabled, to
unlock please press the LOCK button once again.
5.6 Locking the Front Panel
Press the current knob for 3 seconds to lock or unlock the front panel. The current and voltage knob as well as the
SET button are not operational when the front panel is locked.
5.7 Remote Sensing (Optional)
The power supply is optionally equipped with remote sensing terminals, which can read the voltage at the load and
therefore compensate losses. Refer to below figure on how to connect the load to remote sensing terminals. The
remote sensing input can be connected optionally, and is automatically deactivated if not connected.
WARNING: In case you physically disconnect the output leads from the load, also disconnect the remote sensing
leads at the same time.
DSC-Electronics Germany • Georgstraße 36 • 53111 Bonn
DSC-Electronics Germany • Georgstraße 36 • 53111 Bonn