Fault
There is no water coming
out from the handpiece
There is air in the water pipe.
(automatic water supply
mode).
If the problem still can't be solved, please contact with local dealer or manufacturer.
Manufacturer will provide circuit diagrams, component part lists, descriptions, calibration instructions to assist to service
personnel in parts repair.
3.6.2 Notice
[Note 1] The water control knob can adjust the water volume according to the symbol.
[Note 2] To clean the water pipe with the multi-function syringe of the dental unit ( see picture 7):
a) Snip the water pipe at a distance of 10cm to 20cm from the water entrance.
b) Turn on the power switch, get through to the power.
c) Connect the multi-function syringe of the dental unit to the water pipe.
d) Screw off the scaling tip or pull out the handpiece.
e) Step on the foot switch.
f) Turn on the switch of the multi-function syringe, press the air or water into the water pipe to clean and eliminate the impurity.
Possible cause
Multi-function syringe
Picture 7
Solutions
Turn the water control to the Max, reinsert the bottle.