DTF PURENEB AEROSONIC: PURENEB AEROSONIC+ Gebrauchsanweisung Handbuch - Seite 6

Blättern Sie online oder laden Sie pdf Gebrauchsanweisung Handbuch für Produkt für die Atemwege DTF PURENEB AEROSONIC: PURENEB AEROSONIC+ herunter. DTF PURENEB AEROSONIC: PURENEB AEROSONIC+ 12 Seiten.

2.4. Symbols used on devices

Device type BF
Device class 2 (electric)
Caution / Warning
Keep dry
Fragile
3.

PRECAUTIONARY MEASURES

 Aerosol treatments can only be done on medical prescription and within the prescribed doses.
 Carefully keep these instructions for use as it provides necessary information for the good use of the
equipment.
 Before connecting the device to the mains, check the values on the device (volts and Hz) and your
distribution network.
 The warranty does not cover faulty connections.
 Fragile medical device, to be manipulated with care.
 Keep this medical device out of the reach of children or animals (potential risks with the tubes and
detachable part of the nebulizing kit).
 Connect the device to an easily accessible plug, to allow a rapid disconnection from the mains.
 The device has to be installed and put into service according to the information supplied with the
attached documents.
 Before connecting the device to the electric power supply, make sure that the nebulizing kit is
connected to the unit in order to avoid an abnormal noise.
 If the compressor with the nebulizing kit is noisy, do not hesitate to contact your supplier.
 Do not exceed 6 ml medicine in the nebulizing kit EXTRANEB+ MINIDOSE (NL9M) and EXTRANEB SONIC
NASAL (NL11SN).
 Do not exceed the mark "full" printed on the nebulizing kits EXTRANEB and EXTRANEB SONIC.
 Clean if necessary the jets of the nebulizer with hot water, without using any metal wire.
 A modification of jets dimensions can have a bad effect on the particle size and on the therapeutic
aerosols efficiency.
 Each nebulizing kit is intended for a single patient.
 Caution: before using the device, make sure that the nebulizing kit is well mounted.
Check the jet is not clogged.
 Caution: This medical device is directly connected to the electric power supply, do not use it in bath or
under the shower, do not immerse.
 Keep this medical device away from any high heat source.
 Store properly the electric cord and the tubes after each use.
 Pureneb devices are designed to operate at altitudes less than or equal to 3 000 meters.
 No special knowledge or training is required to use Pureneb devices.
 The electric cord cannot be replaced by the maintenance staff.
 An animal can eat the electric cord. Thoroughly check the state of the equipment before use.
See instructions for use
Serial number
Keep away from light
IMPORTANT REMARKS
Alternative current
O
I
Off
Classic Modus
6/12
Manufacturer
Dangerous
voltage
Device reference
II
Sonic Modus