8.4 Stability Filter

In adverse environmental conditions, or when scale is being used in unstable weighing environments, you can
increase the weighing stability with filters. To access this function, press
Press
until display shows "FiLTEr". Press
accordance with the following table:
Filter 0 - Factory programmed (Recommended)
Filter 1 - Stable environments (fast response)
Filter 2 - Not really stable environments (medium response speed)
Filter 3 - Instable environments (low response speed)
Filter 4 - Very unstable environments (slow response speed)
Filter 5 - Strongly unstable environments (very slow response speed)
Filter 6 - Extremely unstable environments (very slow response speed)
To change, press
or

8.5 Sound activation

Keypad produces a sound when pressing keys. This sound can be disabled, press
until display shows "Beep" on display.
Press
to confirm, use
0-
Sound off
1-
Sound on

8.6 AD converter verification function.

This function shows the value of analogical/digital converter to verify the proper function of the entire scale
including load cells. It is useful for our technician to check and identify eventual scale failures.
Press
, then press
number which normally changes.
To escape this function, press
EB300 User Manual
to confirm, and display will show a digit on the right, in
, confirm by pressing
to select the right digit in accordance with the following options:
until it shows: "i count". Press
and then press
and display will show "PLU XX".
. To escape, press
to confirm, and display will show a 6 digit
.
and then press
Page 15 of 18