6.5 Percentage Setting Function

Press
key to switch to "%" and enter percentage weighing ; if you need
resample, please take away the object from the platform, and press
display"0", then press
percentage value of sample, press
same percentage samples on the centre of platform, press
when the reading value becomes stable, Normal percentage and weighing can
be started on.
NOTE: If the balance displays "no", that indicates the samples' weight is too
light

6.6. Backlight Function

Press
key to sw1tch backlight, "ON" indicates backlight is always on,
"oFF" indicates backlight is always off, and "Auto" indicate backlight is auto

6. 7. Unit Selection

This balance has the function to select these units: g, kg, cL oz,
ozt, lb , I, tl, dwt, gn, pes , % Press
7.
RS-232C Communication
RS232C interface: Connecting to PC or printer
Wiring diagram is as follows
Baud rate= 9600bps, check bit= None, data bit =8Bit,
stop bit=1 Bit, input instruction format please read "Key"
function; output data format: 7 bytes of data code +3 byte unit
code +2 byte end code, all the characters are ASCII code
6 |
ANYLOAD ES-HB Precision Balance Product Manual (V1611)
key to display "PEr=xxx", and "xxx" indicates the
key to change the value , and put the
to switch.
key to
to confirm ,