2.3 AUTO OFF SETUP

When scale is on and at zero and stable,
long press UNIT
off setup, use
ZERO
or disable, then press PRINT
confirm and back to weighing mode.
on
enable (scale will turn off after 10 mins haven't been use; to be able to use the scale, needs to turn
on the on/ff switch)
off
disable

2.4 DIVISION SETUP

When scale is on and at zero and stable, long press
ZERO
key to select the divisions, then press PRINT
-301_Calibration
A. Setting up division, decimal point, max. capacity and calibration
Press and hold "PRINT" key,then turn on the scale,to enter division setting:
1. Division Setting:
LCD shows [d
press "PRINT" key to confirm,and enter the decimal point setting.
Ex. When display shows [d
decimal point setting.
2. Decimal point setting:
LCD shows [p
0.0000,P.00000,and then press "PRINT" key to confirm,and enter the maximum capacity
setting.
key enter the auto
key to enable
key to
x]
then press "ZERO" key to select 1、2、5、10、20、50,and then
5],Press "PRINT" key,the division set as 5,then enter the
x]
then press "ZERO" key to select 0、0.0、0.00、0.000、
FUNCTION
key enter the division setup, use
key to confirm and back to weighing mode.