In case of equipment failure
WARNING
If at any time the TRIOS IOS system malfunctions,
or if you suspect in any way that the TRIOS IOS system
is not working correctly:
• Remove the scanner from contact with the patient .
• Unplug the system and make sure it cannot be used
before it is checked .
• Contact your reseller .
• Do not attempt to open any covers on the TRIOS IOS system .
Portable RF communications equipment
WARNING
Portable RF communications equipment (including periph-
erals such as antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to any part of the
MOVE/MOVE+ cart system, including cables specified by
the manufacturer . Otherwise, degradation of the perfor-
mance of this equipment could result .
Wireless Connection
The wireless scanner is intended to be used indoors, in a domestic
or commercial electromagnetic environment .
CAUTION
Other portable and mobile RF communication equipment
could interfere with the wireless scanner, even if the other
equipment complies with CISPR emission requirements .
CAUTION
The wireless scanner could interfere with other portable
and mobile RF communication equipment .
CAUTION
It is recommended to keep a line of sight between the
wireless scanner and the USB wireless dongle attached
to the PC .
5.2 Mechanical Hazards
Fragile and movable parts
WARNING
Be careful to avoid the following potential sources of injury:
• Do not lean or sit on any part of the system .
• Movable parts can pinch body parts .
Dropped or damaged equipment
WARNING
If you drop a scanner tip on the floor, you MUST dispose
of it immediately and NOT use the same tip again for
scanning . There is risk that the mirror in the tip has become
dislodged and can fall out .
WARNING
If the handheld scanner is dropped or bumped, it should
immediately be calibrated before further use . If calibration
fails, please contact your technical service provider . See this
guide's instructions on calibrating the handheld scanner .
8
WARNING
If the handheld scanner is dropped, examine the TRIOS
IOS system for any mechanical damage . If a mechanical
damage is observed or suspected, treat this as an equip-
ment failure (see this guide's instructions on handling
equipment failures) .
WARNING
If a battery is dropped, examine it for any mechanical
damage . If a mechanical damage is observed or suspected,
the battery must be disposed of in the correct manner .
System S3A (MOVE/MOVE+ cart) only
WARNING
• Keep both hands on the upper part of the MOVE/MOVE+
cart when moving it across the floor .
• Do not attempt to lift the MOVE/MOVE+ cart alone . When
moving the system over a door step or other unsmooth sur-
face, get assistance from another person to ensure stability .
• When moving the MOVE/MOVE+ cart across the floor,
maintain all its wheels in contact with the floor and ensure
that they are not locked .
• Always place the scanner in its intended mount on the
MOVE/MOVE+ cart when not in use and when moving
the MOVE/MOVE+ cart .
• Keep your feet clear of the MOVE/MOVE+ cart's base
and wheels when moving the MOVE/MOVE+ cart .
• Never drag the MOVE/MOVE+ cart across any of its cables .
• Do not place the MOVE/MOVE+ cart on a slope greater
than 10 degrees .
• Keep sharp objects away from the touch screen .
System S3P (pod) only
WARNING
• Always place the scanner in its intended mount on the
pod when not in use .
• Do not place the pod on a slope greater than 5 degrees .
5.3 Explosion Hazards
Environment
WARNING
The TRIOS IOS system is not designed to be used in poten-
tially explosive environments, such as in close proximity
to flammable liquids or gases, or in oxygen-enriched
atmospheres .
Distance to the patient with anesthetic
WARNING
There is a potential explosion hazard if the TRIOS IOS
system is operated in the presence of flammable anesthetic .