Customising Settings
There are several settings on the CellAir
o
o
Temperature
C/
F
PPM
CO
2
c
1. Press the
button to switch between
Using the "DIAG" function
PPM
CO
2
c
DIAG
1. Press the
button until "DIAG" flashes
2. Press
, the four LED's on the SEU will flash and
its alarm will sound
3. The four LED's on the RDU will flash and its alarm
will sound
Using the "CALI" function
PPM
CO
2
c
CALI
1.
Press
button until "CALI" flashes
2.
Press
, the "CALI" icon shown on the LCD
3.
Press and hold for at least 10 seconds. The "CALI"
icon will flash and the unit will calibrate
automatically. After 10 minutes the LCD will
display "Pass" or "Fail". If "Fail" appears please try
calibrating the unit again.
Note: If possible, calibrate the SEU outside in ambient
air. Wait 10 minutes for the unit to stabilise before
calibrating. DO NOT breathe on or near the unit during
calibration.
Installation Instructions
1.
Choose a suitable location to install the SEU. Fix the panel holder on the wall at the recommended height of
about 0.45m (1.5 feet) from the floor and as close to the manifolds and valves as possible.
2.
Put the SEU on the panel holder, making sure the connection is tight.
3.
Fix another panel holder in a suitable location outside the monitored space. Push the RDU onto the panel
holder and stick the warning paper next to the RDU.
4.
The communication cable is pre-wired to the SEU. Route this through to the RDU and plug it into the input
port. Communication between the two units is now established.
5.
The CellAir
has one relay output which is pre-wired to the SEU. The relay can be used to operate an external
TM
alarm system or ventilation system. It will be triggered when the CO
level.
6.
After completing the installation, remove the rubber cap from the gas entry point, power the unit up and use
the "DIAG" function to verify the connection between the SEU and the RDU.
w: www.duomo.co.uk
-
that can be altered to personalise it for your application.
TM
PPM
CO
2
F
o
C and
PPM
CO
2
c
DIAG
PPM
CO
2
c
CALI
-
Setting the first alarm parameter
o
F
1. Press
button until "AL1" flashes
2. Press
, the "AL1" icon shown on the LCD
3. Press
to go through the parameters
4. Press
again to save the setting
Setting the second alarm parameter
1. Press
button until "AL2" flashes
2. Press
, the "AL2" icon shown on the LCD
3. Press
to go through the parameters
4. Press
again to save the setting
Note: The second alarm level should be higher than
the first alarm level.
Using the "ReFactSet" function
PPM
CO
2
ReFactSet
c
1. Press
button until "ReFactSet" flashes
2. Press
, then
3. Press
again to save the setting
Only use this function if you wish to return the sensor
to its factory default settings, or if the unit displays
"ESC" on the LCD under conditions of very high
concentrations.
t: 01905 797989
-
PPM
CO
2
c
AL1
AL1
PPM
CO
2
c
AL2
AL2
CO
2
ReFactSet
c
to select either "Yes" or "No"
concentration exceeds the first alarm
2
f: 01905 774296
© Duomo (UK) Ltd. 2007
|
%
PPM
CO
2
c
%
PPM
CO
2
c
CO
2
ReFactSet
c