Duomo SGF 104 Installations- und Benutzerhandbuch - Seite 4

Blättern Sie online oder laden Sie pdf Installations- und Benutzerhandbuch für Zubehör Duomo SGF 104 herunter. Duomo SGF 104 8 Seiten. Oxygen sensor

Duomo SGF 104 Installations- und Benutzerhandbuch
Sensor Status indication
Connections and P.C.B. Layout
Switch 1 - Selection of positive
safety
Switch 2 - Operating range.
Data connection tool TS007
Connecting the tester TS1007 it is possible to sensor history and print
the test report. This report will show the information listed below.
Typical print out
ADC 80 16 (125 145)
1
probe sgF104
sn 24ad12
date 15 2014
DaY 2190
cy 5
al 15
drift 12
range 20
lel 20% (fault)
print
tar
new save
low bat
Page 4
Operating LED
R - +
C NC NO
Gas probe SGF104
S.N. ED567D
Year 12/2015
Batch: SGF104/001
Gas : Oxygen
Power supply 12/27VDC
Absorption 110 mA
Relay 1A 30V SELV
Signal 4 - 20 mA
unlock
pivot
Display "ADC" shows the decimal value of the capsule and the type of "GAS"
Displays the type of sensor,
Displays the serial number,
Displays the manufacture date,
Displays the remaining days at the end of the operating cycle.
Displays how many times the probe has been turned on.
Displays how many times is gone into alarm
Displays how many corrections have been done
Displays the operating range if at 20 or at 100% of LEL
Displays the percentage of dispersed gas
Displays if the probe is in fault
Indicates if you want to print the ticket
Indicates if you want to reset the probe to the data factory.
Asks to save the modified data
Indicates when the battery is low
Green: Operating
Red: Flashing: detecting gas or Steady: alarm threshold
exceeded.
Yellow: Fault
Detection data memory through
USB port
Connect the sensor to the TS1007
and read the data transmitted.
Legend of display