a Fixed transmitter manufacturers, such as: base stations for wireless (cellular / cordless) phones and ground mobile
radios, amateur radio, AM (Amplitude Modulation) and FM (Frequency Modulation) radio broadcasts, and
television broadcasts. The field strength is theoretically It is impossible to predict accurately. In order to evaluate the
electromagnetic environment of fixed RF transmitters, the survey of electromagnetic sites should be considered. If
the measured field strength of the infrared forehead thermometer is higher than the RF compliance level of the above
application, the infrared forehead thermometer should be observed to verify its normal operation. If abnormal
performance is observed, supplementary measures may be necessary, such as reorienting or positioning the infrared
thermometer.
b In the entire frequency range of 150KHz-80MHz, the plant strength should be lower than 3V / m.
RF conduction
GB / T
17625.6
RF radiation
GB / T
17626.3
3 Vrms
150 KHz to 80 MHz
3V/m
80 MHz to 2,5 GHz
12 / 16
Portable
communications equipment should not be
used closer to any part of the infrared
thermometer
isolation distance, including cables. This
distance should be calculated by a formula
corresponding to the frequency of the
transmitter.
Recommended isolation distance
Not
p
d =1.2
applicable
p
d =1.2
3V/m
p
d =2.3
Among them, p is according to the
transmitter's maximum output rated power
provided by the transmitter manufacturer, in
watts (W), and d is the recommended
isolation distance in meters (m).
The field strength of the fixed RF transmitter
is
determined
electromagnetic field, and it should be lower
than the compliance level in each frequency
range.
and
mobile
than
the
recommended
80 MHz to 800 MHz
800 MHz to 2,5 GHz
by
surveying
RF
the