Aplisens SGE-25S.SMART Benutzerhandbuch - Seite 17
Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Zubehör Aplisens SGE-25S.SMART herunter. Aplisens SGE-25S.SMART 20 Seiten. Smart level probes
Auch für Aplisens SGE-25S.SMART: Benutzerhandbuch (20 seiten)
C
Special version of probe.
Cable with additional
teflon covering
-
PROBE
Connection of SGE-25, SGE-25S and SGE-25C in a three-wire system (0...10V output signal).
Not applicable to Ex version.
PROBE
Connection of SGE-25, SGE-16, SGE-25S and SG-25C in a two-wire system
(4...20mA output signal).
Not applicable to Ex version.
Figure 3. SGE-25, SGE-16,SGE-25C and SGE-25S probes – dimensions and connection method
Red or red/yellow wire
-
Teflon tube
Zero
level
SGE25 and SGE-16 probe
Red
Blue
+
0...10V output
_
Black
Screen
Red
+
_
Black
Screen
15
Black or black/blue wire
∅
Capillary
2x0,5
Supporting handle
(Not applicable to SGE-16)
∅
25(SGE-25)
∅
16(SGE-16)
∅
25
Zero
level
SGE-25C probe
Terminal box PP or UZ-2 protection system.
Recommended for all lengths of cable.
Terminal box PP or UZ-2 protection system
(when cable length exceeds 100m).
Black or black/blue wire
-
∅
25
Zero
level
∅
60
SGE-25S probe
+24 VDC
Power supply/
0...10V
measurement system
_
+
Power supply/
measurement system
24 VDC
_
DTR.SG...04
(ENG)