Apogee SQ-100X Series Benutzerhandbuch - Seite 9
Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Zubehör Apogee SQ-100X Series herunter. Apogee SQ-100X Series 19 Seiten. Quantum sensor
DEPLOYMENT AND INSTALLATION
Mount the sensor to a solid surface with the nylon mounting screw provided to prevent galvanic corrosion. To
accurately measure PPFD incident on a horizontal surface, the sensor must be level. An Apogee Instruments model
AL-100 Leveling Plate is recommended to level the sensor when used on a flat surface or being mounted to
surfaces such as wood. To facilitate mounting on a mast or pipe, the Apogee Instruments model AL-120 Solar
Mounting Bracket with Leveling Plate is recommended.
To minimize azimuth error, the sensor should be mounted with the cable pointing toward true north in the
northern hemisphere or true south in the southern hemisphere. Azimuth error is typically less than 1 %, but it is
easy to minimize by proper cable orientation.
In addition to orienting the cable to point toward the nearest pole, the sensor should also be mounted such that
obstructions (e.g., weather station tripod/tower or other instrumentation) do not shade the sensor. Once
mounted, the blue cap should be removed from the sensor. The blue cap can be used as a protective covering for
the sensor when it is not in use.
Nylon Screw: 10-32x3/8
Model AL-100
Important: Only use the nylon screw provided
when mounting to insulate the non-anodized
threads of the aluminum sensor head from the
base to help prevent galvanic corrosion. For
extended submersion applications, more insulation
may be necessary. Contact Apogee tech support for
details.
Nylon Screw: 10-32x3/8
Model AL-120