Datakom DPR-145 Benutzerhandbuch - Seite 13

Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Relais Datakom DPR-145 herunter. Datakom DPR-145 19 Seiten. Temperature protection relay

DPR-145 User Manual

5.4 MODIFYING PROGRAM PARAMETERS

TUŞ

5.5 PROGRAM PARAMETER LIST

DISPLAY
İŞLEM
The parameter to be modified is
selected with MODE button.
When the parameter name is on
display, hold the MENU button
pressed during 5 seconds. The
parameter value will be displayed.
Adjust the parameter using up/down
arrow buttons.
In order to increase/decrease faster
you can hold arrow buttons pressed.
When MODE button is pressed, the
parameter value is recorded and the
next parameter name is displayed.
DESCRIPTION
PT100 CABLE SECTION ()
PT100 cable section in mm
The factory set value is 0.5mm
PT100 CABLE LENGTH ()
PT100 cable length in meters (m).
The factory set value is 0m.
TRIP RELAY TEST ()
Used to test the TRIP relay. The value of
"1" activates the relay. The value of "0"
deactivates it.
2
.
2
.
-13-
V-3.0
(14.09.2016)
EKRAN
MIN
MAX
0.10
9.99
0
999
0
1