Conviron CMP6050 Installationsanleitung Handbuch - Seite 13

Blättern Sie online oder laden Sie pdf Installationsanleitung Handbuch für Kontrollsysteme Conviron CMP6050 herunter. Conviron CMP6050 18 Seiten. Retrofit for 4000/5000 series controller installation instructions

19.
F
Locate the wire labelled 3 from the new control panel assembly and connect it to the existing
3 terminal. Locate the neutral wire (60hz = white; 50hz = blue) leaving the new control panel
assembly and connect to the existing neutral distribution block in control panel (see drawing
257349).
20.
F
Replace existing sensors and/or add new sensors as required. Chamber type and options will
determine which sensors are required:
i.
All chambers will require at least two NTC temperature sensors: One for main temperature
control in the growth environment, and one for the Independent Temperature Shut Off
controller (ir33). To install:
a. Remove existing temperature sensor(s) from aspirator box
b. Install new sensors in aspirator box and route cables for new temperature sensors
c.
d. Wire new Independent Temperature Shut Off sensor to terminal blocks J5 and JC5 on
ii.
If the chamber is equipped with a humidity sensor, it must be replaced. To install:
a. Remove existing humidity sensor from aspirator box
b. Install new humidity sensor in aspirator box and route cables for new humidity sensor
c.
CONVIRON
identically to previously installed sensor(s)
Wire new chamber temperature sensor to terminal blocks J4 and JC4 on new control
panel assembly 3 pole sensor terminals (Figure 15)
new control panel assembly 3 pole sensor terminals.
TB5
J3
J4
J4
B4
J3
JC4
JC4
BC4
identically to previously installed sensor
Wire new humidity sensor to terminal blocks J1, +V and CG (Figure 16) as indicated in
schematic 228124.
TB5
J24
+V
+V
+V
J2
J1
J1
B1
TB5
CG
Figure 15
BROWN
WHITE
BLUE
Figure 16
CMP 6050 RETROFIT INSTALLATION INSTRUCTIONS
TEMPERATURE
SENSOR
CHAMBER
HUMIDITY
SENSOR
HUMMITTER
ONLY