ADE M70011 Betriebshandbuch - Seite 4

Blättern Sie online oder laden Sie pdf Betriebshandbuch für Skalen ADE M70011 herunter. ADE M70011 9 Seiten. Electronic lifter scale approved class iii

Your ADE Electronic Lifter Scale is a class III approved scale which has been manufactured for medical
use in hospitals, by doctors or medical specialists as well as in rehabilitation centres or nursing homes.
The scale is a precision instrument for medical use.
The weighing result is used as an essential basis for medical application or treatment decided
by doctors or medical personnel in hospitals, nursing homes, rehabilitation- and dialysis
centres.
In order to achieve the best possible results kindly read and follow the Operating Manual in all respects.

Check carton contents:

-
Lifter Scale, completely mounted
-
Operating Manual
Unwrapping and installation of scale at suitable patient lifter
-
Carefully remove all packing material being used to prevent transport damages and store this
material for eventual necessary re-sending for warrantee or repair cases.
-
Remove the suspension yoke of the patient lifter and replace this by using the upper
suspension eye of the scale to combine lifting arm and scale.
-
The suspension yoke of the lifter is not attached to the bottom suspension eye of the scale.
-
An essential part to secure utmost safety is the twisting joint of the patient lifter. The ADE lifter
scale can be used with all leading quality patient lifter systems once ADE has approved the
twisting joints used are granting utmost safety.
-
To grant 100% safety we recommend to use ONLY the ADE twisting joint which can
be purchased as accessory to the lifter scale.

Power supply of the scale

The scale is operated by 4 Mignon batteries (1,5V).
One set of batteries is included.

Use of the scale where approved scales are imperative

The scale corresponds to standards and requirements of the European Directives 90 90/384/EEC and
93/42 EEC.

Connecting additional equipment

Conditions for connecting additional equipment are fulfilled via RS 232.
Metrologic requirements for additional equipment like printer and PC have to be metrologically marked
by M on green background.
3

Installation and use of scale

060100-Rev000-M70011