Active Motif 53081 Handbuch - Seite 9

Blättern Sie online oder laden Sie pdf Handbuch für Medizinische Ausrüstung Active Motif 53081 herunter. Active Motif 53081 15 Seiten. Cooled sonication platform

C. Calibration of the Platform Counter
The counter on the platform needs to be set to an
appropriate "zero" point to enable you to return the
platform to the same height consistently once you
have determined the appropriate height for your
purposes. We suggest that the "zero" point be set at
the bottom of the drive screw, meaning the lowest
point to which the platform can be cranked.
To set the platform to the bottom of the drive screw,
turn the crank clockwise until the platform base no
longer continues downward; then, turn the dial on the
analog counter clockwise to "0000" (right). From this
point on, a one unit increase in the counter equals a
0.1 mm increase in the height of the platform.
Note:
If the platform goes too low, it may
not move upward when you begin to
turn the crank. Gently lift underneath
the platform as you crank to engage
the gears completely.
D. Alignment of the Platform Stand
The platform stand can be adjusted in both the horizontal (x-y) and vertical (z) planes, ensuring
that the probe is centered in the tube and that its tip is at the chosen depth in the sample. The
provided Alignment Tool is used to set an approximate alignment for the tube and probe. It is
included because when a tube is placed in a Tube Cooler, you cannot see the probe tip or its
position relative to the bottom of the tube.
The Alignment Tool is used to make initial settings, which
are refined using a tube in the Tube Cooler. (This prevents
possible damage to the probe that could occur if you
simply raised the platform without knowing how deep in
the tube the probe will go.)
You should have already placed your converter/probe in
the bracket and set the counter to "zero", as described
above. To begin, lower the platform and place the
Alignment Tool in the hole on the stand. To lower the
platform, simply press in the latch located between the
counter and the crank (right). After placing the Alignment
Tool in the stand, move the entire platform up to the top
of the rod, against the Stop Collar.
www.activemotif.com
6