Inspecting Straight Diameters (continued)
7. Tighten the indicator clamp.
8. Remove the IT-5100 from the gage block(s).
Reposition the gage on the gage block(s) to
verify the zero setting.
Note: Note the position of the small revolution
counter on the indicator before removing
the gage. Place a piece of masking tape
on the backside of the indicator and record
the dial setting of the small revolution
counter to eliminate incorrect indicator
readings.
9. Remove the gage block(s) from the IT-5100.
10. Set a frequency for verifying the zero setting
of all gages. As a minimum, the IT-5100 gage
should be zeroed on gage block(s) once
during each shift to ensure accurate readings.
11. Press the retraction lever and insert the
IT-5100 gage inside the part.