MAINTENANCE

At least once a week lubricate the O-Ring seals found in the handpiece coupling with silicone lubricating grease. Wearing disposable gloves, apply a
film of grease on the fingertips and lubricate with the fingers.
If liquids that potentially leave incrustations are used to supply the spray, it is recommended that the circuit be flushed well with water at the end of each
use.
The manufacturer requires an annual check or overhaul of the device by authorised service assistance personnel.

5.1. REPAIRS AND OVERHAULS

Installation, overhaul, calibration and repair of the device must be performed exclusively by technical service assistance personnel authorised by the
manufacturer.

5.2. TROUBLESHOOTING

FAULTS
The motor does not turn.
The motor runs but its speed is not adjustable.
The motor functions sporadically.
The motor does not provide sufficient torque.
The motor overheats.
The motor turns regularly.
The burr does not turn.
The handpiece rotates on its coupling.
The handpiece does not remain inserted on
the micro-motor.
Water leakage at the handpiece.
Water leakage at the instrument tubing
coupling.
I-MMr • i-MMr L • i-MMs
PROBABLE CAUSES
Motor blocked.
Blocked handpiece.
Electronic control board broken.
Instrument tubing broken.
Adjustment command broken.
Electronic control board malfunction.
Instrument tubing defective.
Bearings damaged.
Torque set to its minimum.
Endodontics mode selected.
Power supply / motor malfunction.
Insufficient cooling air.
Handpiece with strong friction.
Handpiece not connected properly.
Handpiece broken.
Defective transmission coupling.
Blocked handpiece.
Handpiece damaged.
Motor hook ring damaged.
O-ring seals worn or damaged.
O-ring seals worn or damaged.
INTERVENTION SUGGESTED
Request technical service assistance.
Lubricate the handpiece.
Send the handpiece to technical service
assistance.
Request technical service assistance.
Request technical service assistance.
Request technical service assistance.
Request technical service assistance.
Substitute the instrument tubing.
Request technical service assistance.
Recover the proper adjustment.
Recover the proper adjustment.
Request technical service assistance.
Request technical service assistance.
Lubricate the handpiece.
Send the handpiece to technical service
assistance.
Slide the handpiece out and reinsert it
properly.
Send the handpiece to technical service
assistance.
Request technical service assistance.
Send the handpiece to technical service
assistance.
Substitute the handpiece.
Request technical service assistance.
Substitute the O-rings.
Substitute the O-rings.