Dynaflow MK5 Eco Handbuch - Seite 5
Blättern Sie online oder laden Sie pdf Handbuch für Bedienfeld Dynaflow MK5 Eco herunter. Dynaflow MK5 Eco 8 Seiten.
Optional Features
Remote Start
When installed, allows the remote starting of a one or more fume cupboards from a remote
location.
Remote Alarm
When installed, allows the remote alarming of one or more fume cupboards. This alarm will result in
the isolation of electrical and gas services.
Remote Isolator
When Installed, allows the emergency isolation of one or more fume cupboards from a remote
location.
"GREEN SASH"
The green sash facility is used to allow the sash to be opened to a pre determined height (eg 50%) If
the sash opening exceeds the set opening for a pre determined time, a WARNING will appear on the
screen asking the operator to close the sash. Not only does this feature prompt the user the keep
the sash in a safe position, but reduces the amount of exhaust air being extracted from the Fume
Cupboard, thus reducing the huge energy consumption required in the laboratory make up air
system.
Scrubber PH Monitoring
When enabled the MK5 controller monitors external PH input via 4-20ma current loop.
Scrubber System
OPTION: (The MK5 Eco Panel has available 40 programmable functions
available that can be turned "OFF" or "ON")
Waste treatment
When fitted, allows control and monitoring of waste water treatment, PH monitoring and dosing
prior to dumping to waste.
BMS Outputs
When installed, allows interfacing to building management systems.
Remote Panel
Used in dual sided fume cupboards allow dual controls over the fume cupboard.
Duct Pressure monitoring
When installed, allow for the monitoring of exhaust duct airflow pressure. It is then used to control
fan drive systems and damper networks.