calculated or referenced from the sensor's catalog or calibration
document. Continue entering the information, as directed, until the
last setup page is reached where Exit is displayed. Press Exit to
return to the Calibration Menu.
11. From the Calibration Menu, press the Two-Point Calibration
(Channels 1 thru 4) button. Select the 5D conditioning module box
wherein data from Step 10 was entered. The reading at the top right
of the display page should increase (or decrease) following the
sensor input. If not, double-check the wiring and setup of the module
parameters. After verifying that the channel data is reading
appropriately, the user may perform a two-point calibration of the
input following the instructions displayed. If not, press OK > Exit to
return to the Home page.
12. From the Home page, press the Meter icon to monitor the Analog
Input values for proper operation and calibration.
13. To set up communications between the unit and an external
computer, install the UtiliPAC410 software program bundled as a
standard accessory with the purchase of Model 4500. To install,
insert the UtiliPAC410 CD in the computer disc drive and double-
click the SETUP program icon. Once installation is complete, locate
and launch the program. When launched, the Serial Communications
Setup page appears. UtiliPAC410 will query the user for the proper
communications port and protocol. The user only needs to change
the COM port to match the one that is currently in use (protocols are
correct) and then press OK. This will test the communications port
between the 4500 instrument and the external computer.