Dynalabs DYN-P-4000 Produkthandbuch - Seite 7

Blättern Sie online oder laden Sie pdf Produkthandbuch für Messgeräte Dynalabs DYN-P-4000 herunter. Dynalabs DYN-P-4000 8 Seiten.

All rights reserved. Reproduction or issue to third parties in any form whatsoever is not permitted without written authority from the proprietors.

3) Operation and Installation

3.1) General

The general sensor connector configuration is given below;
Cable Code/Pin Configuration:
• Red
: V +
Power supply voltage +10 to +40 VDC
• Black
: Ground
Power GND
•X-Axis: Yellow : Signal (+)
Positive, analog output voltage signal for differential mode.
Blue : Signal (-)
Negative, analog output voltage signal for differential mode
WARNING
Never connect the power supply and/or the power ground to yellow, and/or blue
cables.
Never connect the power supply to the power ground. Always use a clean power
source and check the voltage range.
4)

Sensor Static Calibration Verification

Using gravity, voltage values are measured in the + and - directions, providing a value of 1 g.
The measurement should be made as follows;
When the sensitivity value of 4000 series sensors is used with the data
acquisition system, the sensor shows +1 g with the effect of gravity in the
direction of the arrow sign.
When the sensor is positioned in the opposite direction to the axis to be
calibrated, the arrow shows -1g as shown below under the effect of gravity.
Using gravity, the voltage values that provide 1 g in the + and - directions are
measured and compared with the catalog value. The calibration value should be
close to the catalog value with a 10% tolerance. Sensor catalog sensitivity
values are given in Table 1.
7