Gait up Physilog 6S Benutzerhandbuch - Seite 9

Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Medizinische Ausrüstung Gait up Physilog 6S herunter. Gait up Physilog 6S 18 Seiten.
Auch für Gait up Physilog 6S: Schnellstart-Handbuch (6 seiten)

⋅ Click on "Done" and disconnect phone from computer
⋅ Open the nRF Connect app on the iOS device and enable Bluetooth
⋅ Open DFU option of the app
⋅ Click on "Select File" and go to the tab "user files" where you find the firmware added to the
app earlier
⋅ Gently shake the Physilog 6S to enter detection mode
⋅ Click on "Select device" button and look for surrounding Physilog
⋅ Select connect of one Physilog 6S sensors in the list
⋅ Plug the selected Physilog to a power source to be sure to have enough battery to perform
the update
⋅ Click on "Upload" and wait until the percentage has arrived at 100%, this may take some
minutes
⋅ Plug the Physilog® to the computer and update the date and time (see above)

5.4. Tips for best practices

Physilog 6S should be handled carefully. In particular, it should not receive shocks, such as fall,
crushing, being hit etc. Do not press too hard on the on/off button, this can damage the button.
Warning: Do not use or charge the Physilog® if the sensor or its case is damaged.
Contact Gait Up about what further action is needed when the case is damaged. In
case the battery is damaged, immediately move the device away from flammable
materials and contact Gait Up's customer support.
⋅ To benefit from the maximal battery life, fully charge Physilog® sensors before doing
measurements (LED shines green when connected to computer or charger, you can also
check with the App).
⋅ Copy trial recordings to your computer: Physilog® is not the best place to keep important
data.
⋅ Physilog 6S can be used outside when the temperature is not exceeding 30 °C. Do not use
Physilog 6S to measure water-sport activities without additional waterproofing.
⋅ Do not introduce pointy objects into the small hole on the bottom of the sensor, this
irreversibly damages the waterproofing of the sensor.
⋅ Sensors do not need systematic yearly recalibration. Depending on the use, regular
recalibration can be recommendable for: offset and gain dependent measures such as
sensor orientation calculation.
⋅ Make sure to know which Physilog® file corresponds to which trial for your analysis – First
digits of file name are automatically incremented for each day of measurement (exceeding
99 resets the increment back to 1). It is recommended to delete old files from the sensor
memory. Note that date of file generation (recorded between one start and stop of the
Physilog 6S user manual v1.0.0 Date: 08.01.2021
Copyright © 2021, All Rights Reserved - Gait Up SA. CH-1015 Switzerland
www.gaitup.com
9