Briefly, the first mode one is called "Physilog Measure" and allows to start and stop data recording
of the Physilog® of up to two sensors at the same time. When pushing the "Start measurement"
button, the connected Physilog start recording and sending the data file to the tablet. When
launching two Physilog®5 at the same time using the tablet application they can be synchronized in
post-processing in Matlab.
The data is streamed to the tablet during the measurement. For high sampling frequencies (over
64Hz) and multiple Physilog®, the data transfer encounters some delay and remaining data is
transferred after the end of the measurement.
Data saved on the tablet only contains accelerometer and gyroscope data, even if the barometer
was turned on. The original data with all sensors is saved on the Physilog®.
During the measurement, the Physilog®5 blinks green where each blink is made of two short flashes.
To stop the measurement, click on the "Stop measurement" button in the application. In case the
sensors have not sent all data to the tablet during the measurement, they twinkle orange to show
that data transfer is still ongoing. The tablet application shows the percentage of the file already
transferred.
The second mode is called "Physilog Visualize", it can be used to check if the sensor's data looks as
expected. Up to 4 sensors can be connected to the application at the same time for this mode. The
"Physilog Visualize" starts displaying data from connected sensors automatically (no start button).
It does not store the data, neither on the tablet nor on the Physilog®. During the measurement, the
Physilog®5 blink green where each blink is made of two short flashes. To stop data streaming either
disconnect the sensor or exit this part of the application.
Note: Bluetooth 4.0 or higher is required on the mobile device to be able to communicate with
Physilog®5.
Physilog5 user manual v1.2.6 Date: 19.06.2018
Copyright © 2018, All Rights Reserved - Gait Up SA. CH-1015 Switzerland