3.2 System Installation
This section thoroughly describes how to install the W8 Physical Cytometer with a straightforward procedure that
requires few simple steps. Please refer to the W8 Physical Cytometer User Manual for any further information.
Step 1
First, define the area of use and the position of the instrument.
• The W8 Physical Cytometer is designed to be positioned onto a fixed laboratory benchtop.
• Its highest performance can be reached when placed onto an anti-vibration support (i.e.
Atomic Force Microscopy benchtop).
• The device should not be located onto a normal table or any surface or support in close
proximity of a source of vibration (i.e. pumps, sonic baths etc..).
• In the case of analyzing or operating with hazardous or unknown materials/substances,
the instrument must be placed and used under a Class II / III Biological Safety Cabinet.
• Do not use the instrument in a wet location or environment. Particularly, do not place the
instrument on areas with high humidity.
• Maintain at least 15 cm of distance between the W8 Physical Cytometer and any wall, object or
PC. This distance is also needed for proper ventilation.Moreover, keep a free area in front of the
instrument to allow the opening of the safety door for proper use (Fig. 3). The main switch button
of the instrument and the plug of the power cable must be easily accessible to the operator.
This is crucial in case of emergency when the instrument needs to be turned off and the power source
needs to be unplugged.
Fig. 3. Schematic representation of the positioning. Distances from walls, objects and/or PC for safety and proper
ventilation.
Step 2
Susbsequently proceed as follow:
• Connect the power source supplier with its cable (keep the cable unplugged from the main
power source at this stage).
• Connect the power supply cable to the power supply plug on the rear of the instrument.