iHealth ABP100 Benutzerhandbuch - Seite 16
Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Blutdruckmessgerät iHealth ABP100 herunter. iHealth ABP100 16 Seiten. Ambulatory blood pressure monitor
870
930
1720
1700-1990
1845
1970
2450
2400-2570
5240
5100-5800
5500
5785
Phenomenon
Electrical
fast
transients/burst
Surges
Line-to-line
Surges
Line-to-ground
Conducted
disturbances
induced by RF fields
Voltage dips
Voltage interruptions IEC 61000-4-11
Date of issue:Sep25,2017
ABP100-SMSY01 V1.0
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 9V/m
Table 4 – Input a.c. power Port
Basic
EMC
standard
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
16 / 16
Immunity test levels
Home Healthcare Environment
±2 kV
100kHz repetition frequency
±0.5 kV, ±1 kV
±0.5 kV, ±1 kV, ±2 kV
3V, 0.15MHz-80MHz
6V in ISM and amateur radio bands between
0.15MHz and 80MHz
80%AM at 1kHz
0% U
; 0.5 cycle
T
At 0º, 45º, 90º, 135º, 180º, 225º, 270º and 315º
0% U
; 1 cycle
T
and
70% U
; 25/30 cycles
T
Single phase: at 0º
0% U
250/300 cycles
T;