Principles Of Operation 1 
™ 
The MR‐100
Plus    Manual Resuscitator consists of four major components 
(figure 1): non rebreathing valve assembly (A), silicone bag (B), intake valve 
(C), and reservoir (D), Part D should be removed if supplemental oxygen is 
not to be supplied from an external gas source. 
TM
Figure 1. The MR100
Plus    Manual Resuscitator 
 
Gas is delivered to the patient by squeezing the bag (see figure 2).    Positive 
pressure within the bag caused by its compression closes the intake valve (E) 
located at the base of the bag. This causes the duckbill valve (F) to close off 
the expiration ports(G) of the non‐rebreathing assembly. Further 
compression of the bag forces gas through duckbill valve to the patient. If 
supplemental gas is used, oxygen is delivered to the reservoir (D) during 
compression of the silicone bag. 
Figure 2. Exhalation 
 
 
 
Principles Of Operation 2 
Exhalation begins when the patient exerts positive pressure (during passive 
exhalation) on the patient side of the duckbill valve (F) or when the operator 
releases pressure from the silicone bag (8). The valve lifts directing the 
patient's expiratory gases through the expiration ports(G) of the non‐ 
rebreathing valve (see figure 3). 
Figure 3. lnspiration 
 
The silicone bag refills for the next breath during patient exhalation.   
Negative pressure within the bag (caused by the expansion of the 
compressed bag) opens the intake valve (E), allowing gas to enter the bag 
either from the atmosphere or from the reservoir (D). 
 
The reservoir should be used whenever supplemental oxygen Is delivered.   
(Supplemental oxygen may be administered without using the reservoir but 
maximum available oxygen concentration will be reduced.)    Excess oxygen 
vents to the atmosphere through the safety valve 
(H) should the reservoir fill before the next delivered breath. If the volume of 
gas in the reservoir is inadequate to fill the silicone bag, room air may be 
drawn in through the reservoir valve safety inlet (I). The concentration of the 
oxygen‐enriched gas entering from the reservoir will depend on factors such 
as oxygen now rate, tidal volume, ventilation rate and operator technique.