Apply an even layer of grease, ap-
Transducer Cable
proximately 1/8" [3 mm] thick to the
flat surface of the transducer. Place
the transducer on the prepared area
of the pipe, with the cable pointing
downstream. See Figure 3. Align the
transducer with the pipe, ensuring
that it is parallel with the pipe wall.
Apply only enough pressure to hold
FLOW
the transducer in place. If no reading
is shown, perform a "rub test" by rub-
bing thumb across the dry transducer
head. The zeros should display a low
Figure 3
flow reading. See Troubleshooting
Guide for further recommendations.
Figure 3
F. OBTAINING A READING
Press the ON/OFF key on the keypad. On power up, the UFX performs
internal diagnostics and starts the ultrasonic transmitter. The display
will read Hold for approximately 5 seconds during this time. If the
transducer is properly mounted to a pipe containing flowing liquid with
at least 100 ppm of 100 micron or larger particles or entrained gases,
the display will begin to indicate velocity. The default units of measure
are FPS (feet per second). The response time of the UFX is approxi-
mately 10 seconds to ob-
tain a reading. Adequate
signal strength is obtained
when the OK icon appears
in the upper right corner of
the display. See Figure 4.
Readings taken without
Figure 4
this icon present may be
erroneous.
It is recom-
mended that survey read-
ings of at least 30 seconds be taken to ensure reading stability.
The UFX can be turned off by pressing the ON/OFF key on the keypad,
or the unit will automatically turn off after approximately 3 minutes of
operation.
G. CHANGING UNITS OF MEASURE
Toggle between FPS (feet per second) and MPS (meters per second)
by depressing the FT/SEC M/SEC key on the keypad.
10