At the same time, the Inhale Valve (G) is opened by the expiratory pressure created by releasing the bag, 
and send the air into the bag through the top of the Reservoir Valve, and at the same time, send the 
Oxygen into the bag from the Oxygen Reservoir till the bag returns to the original shape before 
compressing. To avoid excessive Oxygen flow rate and low compressing frequency causing too high 
pressure inside the bag and the Reservoir, the Reservoir Valve (I) is specially designed to release the 
excessive air, to keep a low‐rate Oxygen supply and ensure the patient's safety 
 
Principle Drawing (D) 
GaleMed reusable resuscitators are equipped with pressure relief valves (optional), automatically provide 
and adjust the pressure in the lung, and keep it within 40 / 60 cmH
will cause the pressure relief valve to jump off and push the pressure out to ensure the patient's safety. 
 
 
Should higher inspiratory pressure be required the pressure relief valve may be overridden by placing the 
thumb ever the valve as show as follows. 
 
3.1.8 Direction for Use: 
3.1.8.1 Reusable‐Resuscitator Operation 
 
1. Place the patient on back, pull his chin upward as 
 
possible to keep the airway and the mouth cavity in 
 
alliance line, so the patient can breathe smoothly.   
 
 
 
2. Clean all visible foreign material inside the mouth and 
 
the throat.   
 
 
 
 
3. Insert the oropharyngeal tube, keep the patient's mouth 
 
open to prevent tongue from occluding the airway. (Can 
 
use a mouth opener to open his mouth) The 
 
oropharyngeal tube can be selected according to the 
 
patient's mouth cavity size.   
 
 
 
4. The emergency personnel should stay behind the 
patient's head, extend the head back and pull his chin 
upwards and towards the emergency personnel. 
Pressure 
Relief Valve 
Pressure Relief Valve Motion 
 
Pressure 
Relief Valve 
Overriding the Pressure Relief Valve 
O, any pressure exceed this standard 
2