Remark: If the patient already has an airway inner tube
inserted, or has been through an airway excise resect
operation, then please remove the mask, connect the
Non‐rebreathing Valve connector with the airway inner
tube, then following the standard operating instruction.
5.
Cover the patient's mouth and nose with the mask, and press
palm against the mask to keep it close to the patient's face.
6.
Use the other hand to press on the Resuscitator, regularly
compress sending with sufficient inhale/exhale frequency.
(Adult: 12‐16 times, Child: 14‐20 times, Infant: 35‐40 times)
7.
The emergency personnel should check: to ensure that the
patient is ventilating properly.
•
Observe rise and fall of the patient's chest (accordingly
with the pressing on the Resuscitator).
•
Check the patients lips and face color through the
transparent part of the mask.
•
Check that the patient valve is working properly through
the transparent housing.
•
During exhalation, check that the interior of the mask is
‐ Testing the Resuscitator
GaleMed reusable resuscitator should be tested as follows:
• When first using the new Resuscitator
• (After cleaning and sterilizing)
• After any new parts have been fitted
• Monthly, if the Resuscitator is not frequently used. Equipment required: Test lung, 0‐100 cmH2O
manometer (for Infant and Child resuscitators only), flow meter, regulated gas supply, gas supply tubing.
3.1.8.2 PEEP valve Operation
1. Fit the Flow Diverter over the downstream housing on the resuscitator as shown in the
illustration. Be sure the diverter is fully seated.
2. Position the Flow Diverter as desired to direct exhaled gases away from the rescuer