AGS Carbon Monoxide Installation, Betrieb und Wartung - Seite 11
Blättern Sie online oder laden Sie pdf Installation, Betrieb und Wartung für Gas-Detektoren AGS Carbon Monoxide herunter. AGS Carbon Monoxide 12 Seiten. Parksafe gas detector
Installation, Operation & Maintenance
STEP 2.
•
Ensure you have the correct gas for the device type prior to application.
•
Screw the regulator/valve into the gas cylinder outlet.
•
Once sealed, the regulator pressure gauge (if available) will indicate cylinder pressure.
•
Offer up the applicator hose/cone and fully cover the sensing element on the front case.
•
Open the valve/regulator to allow the gas to be delivered at a pre-set flow rate.
•
Apply gas.
•
The detector Alarm LED will illuminate at the alarm set point.
•
Remove applicator hose/ cone and turn the gas cylinder regulator/valve off.
•
Wait for detector to stabilise / return to normal.
•
Reset the service reminder by shorting out the unmarked header.
•
Test complete.
Always remove the regulator/valve after use - All cylinders will re-seal upon removal of the regulator!
Never use expired or 'near empty' gas cylinders!
Always give at least five (5) minutes between testing the same unit or until gas has fully dispersed!
Always consider safety and use equipment in accordance with Safety Data Sheets!
Specification
General
Product:
Model(s):
Use:
Indicators (LED)
Orientation
Electrical
Max. Power Consumption
Power Voltage Input Range
Communication
Terminal Wire ratings
End of Life/ Service Battery
Max Chain Length
Construction
Dimensions (H x W x D)
Unit Weight (Approx.)
Nominal Orientation
Housing Material
Environmental
Ingress Protection
Storage Conditions
Sensor Operating Temperature
Sensor Operating Humidity
Sensor Operating Pressure
Compliance
CE
Sensor Specification & Performance
Factory Calibration Conditions
Sensor Type
Measuring
Gas Sensor
Range (ppm)
Carbon
0-10,000
Monoxide (CO)
Nitrogen Dioxide
0-20
(NO
)
2
▲ Rising Alarm ▼Falling alarm
Rev: 11
0421
Park Safe Detector
CO (Carbon Monoxide) NO
Indoor, Car Parks, Garages Non-Hazardous Areas (not to be used in potentially explosive atmospheres)
Green (Power), Yellow (Fault State) & Red (Alarm).
Wall Mounted
15mA Max @ 24Vdc
12-32Vdc (24vdc Nominal)
RS485, Modbus RTU
Copper 18AWG (0.75mm2) Min.
8x screw terminals.
CR2032 - 3V 220mA Radial Lithium Annual service reminder activated upon five (5) hours of continuous
power. EOL activated upon five (5) hours of continuous power.
500 Yards approx
4.92 x 3.15 x 1.38"
1.6oz
Wall Mounted (No Limits)
General purpose ABS plastic
IP43
Dry. Temp: -4 ~ 122°F RH: 15 ~ 90%
-4 ~ 122°F
Continuous 30-80% RH Non-Condensing
Normal Atmospheric Pressure ± 10%
BS EN 50270 / BS EN 61010-1
25° ± 5°C - 77° ± 41°F (40-70% RH)
Electromechanical
Calibration Gas
Response (T90)
120ppm CO
<60s
5ppm NO
<60s
2
*EOL – Expected operational Life – Replace unit after this time.
(Nitrogen Dioxide)
2
Recovery (T10)
(Pre alarm warning)
<50s
<50s
Parksafe Gas Detector
Alarm: 1
Alarm: 2
▲25ppm
▲100ppm
▲0.7ppm
▲2ppm
*EOL
(Years)
5
3
11