• Ensure sample temperature is within 4 °C of chamber
temperature.
4. Starting a Test—
• Before starting a test, set the Volatiles: Yes/No option in
the configuration menu. If your sample has volatiles,
select "Yes" otherwise select "No".
• On the main screen, select the test wizard .
• Following several warnings, begin configuring the test by
providing a name for the test using the arrow keys
and then press . (The name of a previous test will appear
by default.)
• Enter initial moisture content if known (optional). Press to
continue or to go back.
• Setup each stage of the isotherm test. A stage is used to
adjust the isotherm settings for an individual sample (up to
20 stages). Each stage is setup with the isotherm method
(DDI or DVS) and the settings for that portion of the
isotherm. Once a stage is created, it can be modified or
deleted anytime before the test is started.
i.e. For milk powder, Stage 1 could use the DDI method to
ramp quickly from 0.1 to 0.4
DVS method to step from 0.4 to 0.8
increments.
• To setup a new stage, select "new" and press . A screen
with several parameters will appear. If a previous test
has been run, the stages for that test will be listed by
default. These stages can be used as-is or updated with new
settings. If you need to delete a stage, go to the specific
stage and press . Scroll down to the bottom of the stage
settings select "Delete" and press .
• Select the isotherm method as either Dynamic Dew Point
Isotherm (DDI) or Dynamic Vapor Sorption (DVS). DDI
will create a dynamic isotherm and DVS will create a
static or equilibrium isotherm. The other choices will
adjust based on which method is selected.