Decagon AquaLab Schnellstart-Handbuch - Seite 5

Blättern Sie online oder laden Sie pdf Schnellstart-Handbuch für Messgeräte Decagon AquaLab herunter. Decagon AquaLab 6 Seiten. Vapor sorption analyzer

Decagon AquaLab Schnellstart-Handbuch
V S A Q u i c k S t a r t G u i d e
    the stage settings select "Delete" and press  .
  •  Select the isotherm method as either Dynamic Dewpoint
    Isotherm (DDI) or Dynamic Vapor Sorption (DVS). DDI
    will create a dynamic isotherm and DVS will create a
    static or equilibrium isotherm. The other choices will
    adjust based on which method is selected. 
• For DDI method
To choose the starting 
        highlighted and then cycle to the desired value using
        the up and down arrows. Any value between 0.03
a
        and 0.95 
 can be chosen including "current". 
w
        Choosing "current" means that the stage will start at
a
        the initial 
 of the sample. Press   when finished.
w
a
To choose the final 
        highlighted and then cycle to the desired value using
        the up and down arrows. Any value between 0.03
a
        and 0.95 
 can be chosen including current (see
w
        "current" definition above). Selecting a final 
        is lower than the starting 
        while selecting a final 
        will result in adsorption. Press   when finished.
To choose the temperature of the stage, press   with
        "Temperature" highlighted and then cycle to the
        desired temperature using the up and down arrow keys
        (hold down arrows for accelerated scrolling). Any
        temperature between 15-60°C can be chosen. Each
        stage can run at a unique temperature.
To choose the desired 
        "Resolution" highlighted and then cycle to the desired
        value using the up and down arrows. Any value
        between 0.003 and 1.0 
        resolution for DDI determines the target 
        resolution. Setting a higher resolution value will
        make the test faster, but will decrease the 
        resolution. A DDI step of 0.01 
To choose the air flow rate, press   with "Flow"
        highlighted, and then cycle to the desired value using
        the up and down arrows. Any value between 
a
, press   with "Start"
w
, press   with "Final"
w
a
 that
w
a
 will result in desorption,
w
a
 higher than the starting 
w
a
 resolution, press   with
w
a
 can be chosen. The
w
a
w
a
w
a
 is recommended.
w
Sample Cup
Stainless steel cups are
recommended. Plastic cups
absorb moisture over time.
Weighing Pan
Air Cylinder
Base Block Bore
a
w
Figure 4: Sample cup and lower block assembly
•  Figure 4: Sample cup and lower block assembly illustrates
    the proper positioning of the parts in the lower block. The
    air cylinder is placed into the bore of the base block. It
    should slide into the block easily and settle at the bottom
    of the bore. The weigh pan connects to the scale and 
    holds the sample cup. For best results, verify cleanliness of
    all parts.
V S A Q u i c k S t a r t G u i d e
Sample Cup Lid
Only used for sample storage and 
must be removed during tests.