Troubleshooting

Nebuliser fails to operate
Check the plug is correctly inserted into the mains outlet socket. If
this does not solve the problem, then the equipment should be
returned to the manufacturer or authorised service location for
repair.
Nebuliser is unexpectedly loud or produces an unfamiliar noise
Disconnect from the power supply and shake gently to check for
parts loose inside. If so, then return the equipment to the
manufacturer or authorised service location for repair.
Usual dose takes longer than normal to nebulise
First, check the filter is clean. If not, wash in hot water with
detergent, rinse, dry and replace. If the filter does not come
clean, then purchase replacements from your distributor.
If the filter is clean, then check the nebuliser chamber has not
been used more than 750 times. If so, then purchase new ones
from your distributor.
If not, separate the nebuliser chamber into its three parts, wash in
hot water with detergent, rinse, dry and reassemble.
If this does not solve the problem, the nebuliser might need
maintenance from an authorised service location.