5-2. Replacing the Fuse
A fuse is used. It is contained in the fuse holder on rear panel of the machine.
Type : Fast-acting fuse (20 mm long, 5 mm diameter, circular tube type)
Rating :
Replacement procedure:
(1) Turn the power switch off.
(2) Remove power cord from wall outlet and cord inlet on rear panel.
(3) With a flat-tip screwdriver, pull out the fuse holder cap on rear panel.
(4) Replace the blown fuse with a new one with the same amperage.
(5) Push in the fuse holder cap securely.
(6) Connect power cord.
(7) Turn the power switch on and test sealer.
5-3. Cleaning the Sealing Head
The sealing head and clamp can easily be stained by dust or liquid. Frequent cleaning
will offer you the best long-term results.
For proper cleaning, use a dry cloth or a cloth dampened with water. A mild detergent
may be used. Thoroughly dry the sealing head area before applying power. If the
sealing head is wet when RF energy is applied, an arc may appear.
Follow the decontamination procedure of your facility as necessary.
Avoid spraying liquids in the sealing clamp or sealing head area.
How to take off and put on the head protector
To take off, push both sides.
SE45010UMC.0306
Power source voltage
100-120 VAC
220-240 VAC
Fuse amperage
3 Amps
2 Amps
To put on, just push in.